ED 23

Heating oven with mechanical control

The strength of the BINDER ED series heating oven: routine drying and sterilization applications up to 300 °C. Thanks to the gravity convection, thermal processes are highly efficient for this heating oven. A BINDER ED heating oven ensures fast, uniform drying.

Data sheet ED 23

overview

Made in Germany


  • Electronically controlled APT.line™ preheating chamber assuring temperature accuracy and reproducible results
  • Temperature range from 5 °C above ambient temperature to 300 °C
  • Digital temperature setting with an accuracy of one degree
  • DS control with integrated timer 0 to 99 hrs
  • Independent adjustable temperature safety device class 2 (DIN 12880), with visual temperature alarm
  • Adjustable ventilation by means of front ventilation flap slide and rear exhaust Ø 50 mm
  • Optional RS 422 Interface for APT-COM™ Data Control System communication software
  • Units up to 115 liters are stackable
  • 2 chrome-plated racks included
  • ED 23 with space-saving door mechanism
  • BINDER test confirmation
  ED 23
Exterior dimensions
Width (mm) 435
Height (incl. feet) (mm) 495
Depth (mm) 520
Plus door handle, I-panel, connection (mm) 105
Wall clearance, rear (mm) 100
Wall clearance, side (mm) 100
Exhaust duct (outer Ø mm) 52
Steam space volume (l) 36
Number of doors (ea.) 1
 
Interior dimensions
Width (mm) 222
Height (mm) 330
Depth (mm) 300
Interior volume (l) 20
Racks (number standard/max.) 2 / 3
Load per rack (kg) 12
Permitted total load (kg) 25
Weight (empty) (kg) 22
 
Temperature data
Temperature range approx. 5 °C above ambient temperature to (°C) 300
Temperature variation 1)
at 70 °C (± K) 1,5
at 150 °C (± K) 2,5
at 300 °C (± K) 3,8
Temperature fluctuation at 70 °C (± K) 0,3
Warm-up time 2)
to 70 °C (min.) 13
to 150 °C (min.) 24
to 250 °C (min.) 35
Recovery time after doors were open for 30 sec. 2)
at 70 °C (min.) 2,5
at 150 °C (min.) 5
at 300 °C (min.) 8
 
Ventilation data
Ventilation
at 70 °C (x/h) 10
at 150 °C (x/h) 13
at 300 °C (x/h) 17
 
Electrical data
IP protection class acc. to EN 60529 IP 20
Voltage (± 10%) 50 / 60 Hz (V) 230,1N~
Nominal power (kW) 0,8
Energy consumption
at 70 °C (W) 43
at 150 °C (W) 148
at 300 °C (W) 450

1) without a viewing window

2) To 98% of the set value
 

All technical data are specified for units with standard equipment at an ambient temperature of +25 °C and a line voltage fluctuation of ±10%. The temperature data is determined in accordance to factory standard following DIN 12880, respecting the recommended wall clearances of 10% of the height, width and depth of the inner chamber. All figures are typical average values for series devices. We reserve the right to alter technical specifications at any time.

 

  • Numerous access ports

    With silicone plugs for introducing external measuring instruments into the chamber, access ports with 10, 30, 50 mm diameters.

  • Door with window and interior lighting

    For optimal process control in the inner chamber, available for all device sizes.

  • Door lock

    Prevents unauthorized access to the process sequences in the chamber.

  • Calibration certificate & validation

    BINDER can significantly reduce the workload in qualifying and validating devices. Nobody knows our devices as well and has as much experience in certifications as we do.

ED 23
Access ports with silicone plug, 10, 30, 50, 100 mm O
Anti-slip rubber pads for safe stacking (1 set of 4 pieces) O
Switchable audible alarm for over temperature. Adjustable limit on the independent temperature safety device O
Independent temperature safety device class 3.1 (DIN 12880) with optical alarm O
Analog output for temperature 4 - 20 mA with 6-pin DIN socket (output not adjustable) O
Temperature measurement acc. to DIN 12880 (27 measuring points, for 23 liter devices 15 measuring points) at 150 °C or at specified temperature with measuring protocol and certificate O
Calibration certificate, measurement in center of chamber at 150 °C or at specified testing temperature O
Extension to calibration certificate. Each additional measurement at additional measuring point or testing temperature O
Data Logger Kit T 350: For continuous temperature recording of 0 °C to 350 °C. Kit includes 1 data logger, PT 100 sensor with 2 m Teflon extension cable and 1 fixture for mounting to the BINDER unit O
Data Logger Software: Configuration and evaluation software for all BINDER Data Logger Kits, incl. data cable O
Rack, chrome-plated O
Rack, stainless steel O
Perforated shelf, stainless steel O
Door lock O
Door gasket, FKM (Viton) O
Door with window 180 x 180 mm and interior lighting (15 W) O