FD 23

Heating oven with mechanical convection

A BINDER FD series heating oven is always used when fast drying and sterilization is required. Thanks to its fully homogeneous temperature distribution, quick dynamics and powerful fan, this heating oven saves valuable time.

Data sheet FD 23

overview

Made in Germany


  • Electronically controlled APT.line™ preheating chamber assuring temperature accuracy and reproducible results
  • Temperature range from 5 °C above ambient temperature to 300 °C
  • DS control with integrated timer 0 to 99 hrs
  • Digital temperature setting with an accuracy of one degree
  • Independent adjustable temperature safety device class 2 (DIN 12880), with visual temperature alarm
  • Adjustable front ventilation flap slide and rear exhaust Ø 50 mm (1.97 inch)
  • Units up to 115 liters are stackable
  • 2 chrome-plated racks included
  • FD 23 with space-saving door mechanism
  • BINDER test confirmation
  FD 23
Exterior dimensions
Width (mm) 435
Height (incl. feet) (mm) 495
Depth (mm) 520
Plus door handle and connection (mm) 105
Wall clearance, rear (mm) 100
Wall clearance, side (mm) 100
Exhaust duct (outer Ø mm) 52
Steam space volume (l) 36
Number of doors (ea.) 1
 
Interior dimensions
Width (mm) 222
Height (mm) 330
Depth (mm) 277
Interior volume (l) 20
Racks (number standard/max.) 2 / 3
Load per rack (kg) 12
Permitted total load (kg) 25
Weight (empty) (kg) 33
 
Temperature data
Temperature range approx. 5 °C above ambient temperature to (°C) 300
Temperature variation
at 70 °C (± K) 0,8
at 150 °C (± K) 2,2
at 300 °C (± K) 4,3
Temperature fluctuation at 70 °C (± K) 0,3
Warm-up time 1)
to 70 °C (min.) 7
to 150 °C (min.) 22
to 300 °C (min.) 45
Recovery time after doors were open for 30 sec. 1)
at 70 °C (min.) 2
at 150 °C (min.) 4
at 300 °C (min.) 9
 
Ventilation data
Ventilation
at 70 °C (x/h) 59
at 150 °C (x/h) 64
at 300 °C (x/h) 53
 
Electrical data
IP protection class acc. to EN 60529 IP 20
Voltage (± 10%) 50 / 60 Hz (V) 230
Nominal power (kW) 0,8
Energy consumption
at 70 °C (W) 145
at 150 °C (W) 300
at 300 °C (W) 720

1) To 98% of the set value
 

All technical data are specified for units with standard equipment at an ambient temperature of 25 °C and a line voltage fluctuation of ±10%. These average values have been determined according to the BINDER factory standard, respecting the recommended wall clearances of 10% of the height, width and depth of the inner chamber. Differing ambient temperatures and production-related device-specific variances can lead to varying technical data.

Therefore, we recommend individual customer-specific calibration or validation for applications on the limit of the permitted ambient temperature range.

 

  • Numerous access ports

    With silicone plugs for introducing external measuring instruments into the chamber, access ports with 10, 30, 50 mm diameters.

  • Door with window and interior lighting

    For optimal process control in the inner chamber, available for all device sizes.

  • Door lock

    Prevents unauthorized access to the process sequences in the chamber.

  • Calibration certificate & validation

    BINDER can significantly reduce the workload in qualifying and validating devices. Nobody knows our devices as well and has as much experience in certifications as we do.

FD 23
Access ports with silicone plug, 10, 30, 50, 100 mm O
Anti-slip rubber pads (1 set of 4 pieces) for safe stacking (not possible for FD 23) O
Independent adjustable temperature safety device class 3.1 (DIN 12880) with optical alarm O
Analog temperature output, 4 - 20 mA with 6-pin DIN socket (output not adjustable) O
Over temperature alarm, audible, can be switched off. Temperature limit can be set at the independent, adjustable temperature safety device O
Securing elements for additional fastening of racks (1 set of 4 pieces) O
Temperature measurement acc. to DIN 12880 at 150 °C or at specified testing temperature with measuring protocol and certificate O
Factory calibration certificate. Measurement in center of chamber at 150 °C (302 °F) or at specified testing temperature O
Extension to factory calibration certificate. Each additional measurement at additional measuring point or temperature O
Data Logger converter cable RS 232 to USB 2.0 O
Data Logger Kit T 350: For continuous temperature recording of 0 °C to 350 °C. Kit includes 1 data logger, PT 100 sensor with 2 m Teflon extension cable and 1 fixture for mounting to the BINDER unit O
Data Logger Software: Configuration and evaluation software for all BINDER Data Logger Kits, incl. data cable O
Rack, chrome-plated O
Rack, stainless steel O
Shelf, perforated, stainless steel O
Locking door handle with key O
Door gasket, FKM (Viton) O
Door with window 180 x 180 mm and interior lighting (15 W) O